CTP-S-011 · Vacuum extraction

Probe-on vacuum sampling

Uses an installed probe-on adapter and vacuum pump to draw oil from a reservoir through new tubing.

Suitable applications

Use this method when the access point supports the evidence required.

  • Gear systems
  • Hydraulic systems
  • Compressors
  • Large circulating reservoirs
Limitations to control

Recognise what can bias the result.

  • Tube depth determines the sampled zone
  • Vacuum leaks can interrupt flow
  • Reservoir sampling is less dynamic than a live-line point
Before approaching the asset

Materials and preparation.

Use laboratory-specified containers and clean dedicated accessories. Keep the cap, bottle mouth and tubing ends protected throughout the collection.

  • 01Installed probe-on reservoir adapter
  • 02Vacuum pump and bottle adapter
  • 03New clean tubing
  • 04Flushing and sample bottles
Controlled method

Field procedure and evidence protected.

Perform the sequence consistently. If a step cannot be followed, record the deviation because it may explain an apparent trend change.

  1. 01

    Stabilise the operating condition

    Collect during normal representative operation or, where the approved procedure requires shutdown, within 10 minutes. Record cold or non-standard sampling conditions.

  2. 02

    Verify the asset and decision point

    Confirm machine, compartment, lubricant, sample-point identity and the maintenance question before opening the kit.

  3. 03

    Prepare clean equipment and records

    Use the laboratory-specified bottle, labels, clean adapters or new tubing, a separate flushing container and the required PPE. Keep caps and bottle mouths protected.

  4. 04

    Inspect and clean the sampling point

    Remove the dust cover, check for blockage or damage and clean the external point without pushing dirt into the system.

  5. 05

    Confirm probe position

    Verify the installed probe reaches the defined active zone and remains above bottom sludge and below the surface layer.

  6. 06

    Fit new tubing and flushing bottle

    Connect clean tubing through the protected pump adapter without touching either end.

  7. 07

    Vacuum-flush the probe path

    Draw enough oil to clear the probe, tubing and adapter while keeping the bottle upright.

  8. 08

    Draw the sample without moving the probe

    Change to the sample bottle and maintain the same tube and probe geometry throughout filling.

  9. 09

    Fill without contaminating the bottle

    Keep the bottle upright and avoid contact with the bottle mouth. Leave suitable headspace; the source guide uses approximately one-half full above 150 cSt, two-thirds for 68-150 cSt and three-fourths below 68 cSt at 40 C unless the laboratory specifies otherwise.

  10. 10

    Cap, clean and restore

    Seal the bottle immediately, wipe the exterior, replace the port cover and return valves, bypasses and filtration equipment to the approved operating state.

  11. 11

    Document and dispatch

    Record asset and oil hours, top-up, filter or maintenance events, operating abnormalities, point, date and sampler. Package securely and send to the laboratory promptly.

Diagnostic quality control

Interferences, effects and controls.

A result may be analytically correct and still misrepresent the machine. Review these controls before escalating maintenance.

Probe-on vacuum sampling sampling interferences
Interference sourcePossible effect on resultField control
Inadequate flushingOld fluid, settled debris or contamination in the adapter and dead volume can dominate the bottle.Calculate dead volume and flush the required multiple into a separate container before filling the sample bottle.
Changing sample locationA clean downstream point, reservoir surface and active return line represent different particle populations.Use the same labelled point and operating state for each trend sample.
Dirty bottle, cap, tubing or adapterAdded particles, water or residue can create false contamination alarms.Use laboratory-specified containers, protect openings and use new or verified-clean wetted parts.
Settling or particle fly-byLow-velocity zones can under-report settled material while poorly located high-velocity ports can miss the target population.Select a turbulent, representative zone and document whether the objective is machine, filter or reservoir condition.
Safety controls

Complete the task without creating a new hazard.

  • Follow plant, OEM and permit-to-work requirements; the website does not authorise a sampling point or override site procedures.
  • Confirm lubricant temperature and pressure before sampling. Use engineered pressure reduction and thermal protection where required.
  • Wear eye and hand protection compatible with the lubricant and process hazards.
  • Declare hazardous, toxic, flammable, biologically active or process-contaminated samples to the laboratory before shipment.
  • Clean skin exposure promptly and follow site medical procedures for eye contact, ingestion or injection injury.
Source download

Retain the original field reference.

Complete manual, pages 35-37

Download source PDFReturn to all sampling procedures →